Contact an Account ManagerDetermination of primary aromatic amines from polyamide kitchenware, validated by the European Union Reference Laboratory (EURL)
Summary
Type: HPLC
Application #: 125750
Detection:
Conditions: Eluent A: 0.01 mol/L sodium acetate, pH 7.2
Eluent B: methanol
Gradient: 0.0-21.1 % B in 2.50 min, 21.1-65.3 % B in 39.50 min, 65.3-100.0 % B in 1.00 min, 100.0 % B for 7.00 min, 100.0-0.0% B in 3.00 min, 0.0 % B for 6.90 min
Flow rate: 0.3 mL/min
Temperature: 30 °C
Injection volume: 100 µL
Precolumn-backflush (column switching valve):
0.00 min back-flushing off
0.80 min back-flushing on
53.00 min back-flushing off
Authors: Simoneau, C. et al.
Source: JRC Sci. and Tech. Reports, EUR 24815 EN 2011